Search

Tapped Density Tester SVM II

Pedir una oferta
Crear PDF

Idioma:

The SVM II is our latest generation of tapped density tester. With unprecedented flexbility and intelligent features, it makes tapped density testing as easy and reliable as never before.

The 7“ high resolution touch display with a modern user interface focuses the user on the most important task the new SVM II has to fullfill - 100% USP/EP compliant tapped density testing. With the automatic calculation features of the TestAssist, the user is guided step-by-step through the testing process as quickly as possible. It also automatically calculates the Hausner Ratio and compressibility index and prints the reports after every test.

In addition to its ease of use, the SVM II is the most flexible tapped density tester we ever built. The newly designed SwitchPlate system enables 5 different versions of the SVM II, with USP 1, USP 2 or USP 3 in either one or two test station configurations or a combination with both methods. And the best part - they can also be upgraded at a later time point!

SVM II also supports a variety of method and norms: From USP 1 and USP 2 to ASTM B527-15 and DIN EN ISO 787-11:1995 - with the SVM II, users are on the safe side of valid and conform tapped density testing.

Tapped Density Tester SVM II

Tapped Density Tester SVM II

Tapped Density Tester SVM II Accesorios

Tapped Density Tester SVM II Datos técnicos

Peso17 kg
Dimensions (H x W x D)196.5 x 300 x 326 mm
Voltage115/230 V; 50/60 Hz
Fuses1 A
InterfacesUSB A/B, Ethernet, RS232
Protection classI/EN 61140
Protection typeIP 21/IEC 529
controlador7" high resolution touch display
Test stations1 or 2 test stations using 5 different SwitchPlates
Supported methods- USP 1: 300±15 taps/min; tap height 14±2mm
- USP 2: 250±15 taps/min; tap height 3±0.2mm
- USP 3 as described in USP <616>
- EP 2.9.34 Bulk density and tapped density of powders
- USP <1174> Powder Flow
- ASTM B527-15 Tap density of metal powders and compounds
- DIN EN ISO 787-11:1995 General test methods, pigments and fillers
Data storage10 reports (temporarily)
ExportNetwork export via external software (Export manager) as XML / PDF
Ambient temprature during operation+10 °C up to +30 °C
Storage and transport temperature+5 °C up to +40 °C
Relative humidity25-80 % non condensing